ࡱ> ;$&      !"#[@ {bjbj44 }C5ViVi ENRRRRtwtwtwwQQQ8Q|HVwmXh"hhhT\$#Ruvtwv"RRhh@<Rhtwh^Vn>m6 tw"hX 'j7@Q&6lTmƞ\,@"wwRRRR0^tw 6JFw,w,  Data Logger Quick Start Guide April 2003 Edition Part Number 123-45 Table of Contents  TOC \o "1-3" \h \z \t "Comment Text,1"  HYPERLINK \l "_Toc39650193" Introduction  PAGEREF _Toc39650193 \h 5  HYPERLINK \l "_Toc39650194" Warranty  PAGEREF _Toc39650194 \h 5  HYPERLINK \l "_Toc39650195" Limitation of Liability  PAGEREF _Toc39650195 \h 6  HYPERLINK \l "_Toc39650196" Battery Warning  PAGEREF _Toc39650196 \h 6  HYPERLINK \l "_Toc39650197" 1 Quick Start  PAGEREF _Toc39650197 \h 7  HYPERLINK \l "_Toc39650198" 2 Software Installation  PAGEREF _Toc39650198 \h 10  HYPERLINK \l "_Toc39650199" Installing from CD ROM  PAGEREF _Toc39650199 \h 10  HYPERLINK \l "_Toc39650200" 3 Hardware Installation  PAGEREF _Toc39650200 \h 11  HYPERLINK \l "_Toc39650201" CRYOTEMP  PAGEREF _Toc39650201 \h 12  HYPERLINK \l "_Toc39650202" OCTRTD  PAGEREF _Toc39650202 \h 13  HYPERLINK \l "_Toc39650203" OCTTEMP  PAGEREF _Toc39650203 \h 15  HYPERLINK \l "_Toc39650204" QUADRTD  PAGEREF _Toc39650204 \h 17  HYPERLINK \l "_Toc39650205" QUADTEMP  PAGEREF _Toc39650205 \h 19  HYPERLINK \l "_Toc39650206" RTDTEMP101  PAGEREF _Toc39650206 \h 21  HYPERLINK \l "_Toc39650207" RTDTEMP110  PAGEREF _Toc39650207 \h 23  HYPERLINK \l "_Toc39650208" SAFETEMP  PAGEREF _Toc39650208 \h 25  HYPERLINK \l "_Toc39650209" TC110  PAGEREF _Toc39650209 \h 27  HYPERLINK \l "_Toc39650210" TC4000  PAGEREF _Toc39650210 \h 29  HYPERLINK \l "_Toc39650211" TEMP100  PAGEREF _Toc39650211 \h 31  HYPERLINK \l "_Toc39650212" TEMP101  PAGEREF _Toc39650212 \h 32  HYPERLINK \l "_Toc39650213" TEMP110  PAGEREF _Toc39650213 \h 33  HYPERLINK \l "_Toc39650214" TEMP1000  PAGEREF _Toc39650214 \h 34  HYPERLINK \l "_Toc39650215" TEMP1000FP  PAGEREF _Toc39650215 \h 35  HYPERLINK \l "_Toc39650216" TEMP1000IS  PAGEREF _Toc39650216 \h 36  HYPERLINK \l "_Toc39650217" TEMP1000P  PAGEREF _Toc39650217 \h 37  HYPERLINK \l "_Toc39650218" TEMP1000S  PAGEREF _Toc39650218 \h 38  HYPERLINK \l "_Toc39650219" TEMPRETRIEVER  PAGEREF _Toc39650219 \h 39  HYPERLINK \l "_Toc39650220" THERMOVAULT  PAGEREF _Toc39650220 \h 40  HYPERLINK \l "_Toc39650221" MICRORHTEMP  PAGEREF _Toc39650221 \h 42  HYPERLINK \l "_Toc39650222" MICRORHTEMP-EB  PAGEREF _Toc39650222 \h 43  HYPERLINK \l "_Toc39650223" RHTEMP101  PAGEREF _Toc39650223 \h 44  HYPERLINK \l "_Toc39650224" RHTEMP110  PAGEREF _Toc39650224 \h 45  HYPERLINK \l "_Toc39650225" RHTEMP1000  PAGEREF _Toc39650225 \h 46  HYPERLINK \l "_Toc39650226" RHTEMP1000IS  PAGEREF _Toc39650226 \h 47  HYPERLINK \l "_Toc39650227" OCTVOLT  PAGEREF _Toc39650227 \h 48  HYPERLINK \l "_Toc39650228" QUADVOLT  PAGEREF _Toc39650228 \h 50  HYPERLINK \l "_Toc39650229" VOLT101  PAGEREF _Toc39650229 \h 52  HYPERLINK \l "_Toc39650230" VOLT110 100mV  PAGEREF _Toc39650230 \h 54  HYPERLINK \l "_Toc39650231" VOLT110 2.5, 15 and 30 V  PAGEREF _Toc39650231 \h 56  HYPERLINK \l "_Toc39650232" OCTPROCESS  PAGEREF _Toc39650232 \h 58  HYPERLINK \l "_Toc39650233" PROCESS101  PAGEREF _Toc39650233 \h 60  HYPERLINK \l "_Toc39650234" PROCESS110  PAGEREF _Toc39650234 \h 62  HYPERLINK \l "_Toc39650235" QUADPROCESS  PAGEREF _Toc39650235 \h 64  HYPERLINK \l "_Toc39650236" EVENT101  PAGEREF _Toc39650236 \h 66  HYPERLINK \l "_Toc39650237" EVENT110  PAGEREF _Toc39650237 \h 68  HYPERLINK \l "_Toc39650238" PULSE101  PAGEREF _Toc39650238 \h 70  HYPERLINK \l "_Toc39650239" PULSE110  PAGEREF _Toc39650239 \h 72  HYPERLINK \l "_Toc39650240" STATE101  PAGEREF _Toc39650240 \h 74  HYPERLINK \l "_Toc39650241" STATE110  PAGEREF _Toc39650241 \h 76  HYPERLINK \l "_Toc39650242" PRHTEMP101  PAGEREF _Toc39650242 \h 78  HYPERLINK \l "_Toc39650243" PRHTEMP110  PAGEREF _Toc39650243 \h 79  HYPERLINK \l "_Toc39650244" PRTEMP101  PAGEREF _Toc39650244 \h 80  HYPERLINK \l "_Toc39650245" PRTEMP110  PAGEREF _Toc39650245 \h 81  HYPERLINK \l "_Toc39650246" PRTEMP1000  PAGEREF _Toc39650246 \h 82  HYPERLINK \l "_Toc39650247" PRTEMP1000P  PAGEREF _Toc39650247 \h 83  HYPERLINK \l "_Toc39650248" pHTEMP101  PAGEREF _Toc39650248 \h 84  HYPERLINK \l "_Toc39650249" LEVEL1000  PAGEREF _Toc39650249 \h 86  HYPERLINK \l "_Toc39650250" LEVEL2000  PAGEREF _Toc39650250 \h 87  HYPERLINK \l "_Toc39650251" SHOCK101  PAGEREF _Toc39650251 \h 88  HYPERLINK \l "_Toc39650252" SHOCK101-EB  PAGEREF _Toc39650252 \h 90  HYPERLINK \l "_Toc39650253" ULTRASHOCK  PAGEREF _Toc39650253 \h 92  HYPERLINK \l "_Toc39650254" PHONEPROCESS  PAGEREF _Toc39650254 \h 94  HYPERLINK \l "_Toc39650255" PHONETEMP  PAGEREF _Toc39650255 \h 96  HYPERLINK \l "_Toc39650256" IROCTTEMP  PAGEREF _Toc39650256 \h 97  HYPERLINK \l "_Toc39650257" IRQUADTEMP  PAGEREF _Toc39650257 \h 99  HYPERLINK \l "_Toc39650258" IRTC110  PAGEREF _Toc39650258 \h 101  HYPERLINK \l "_Toc39650259" IRTC4000  PAGEREF _Toc39650259 \h 103  HYPERLINK \l "_Toc39650260" BRIDGE110  PAGEREF _Toc39650260 \h 105  HYPERLINK \l "_Toc39650261" ENGINEERING UNITS  PAGEREF _Toc39650261 \h 106  HYPERLINK \l "_Toc39650262" Calibration  PAGEREF _Toc39650262 \h 108  HYPERLINK \l "_Toc39650263" NET101  PAGEREF _Toc39650263 \h 110  HYPERLINK \l "_Toc39650264" RFEXT  PAGEREF _Toc39650264 \h 111  HYPERLINK \l "_Toc39650265" MadgeTech Return Policy  PAGEREF _Toc39650265 \h 113  Introduction Small, simple and affordable, MadgeTech data loggers measure and record data at user specified intervals ranging from once every 2 seconds to once every 12 hours. MadgeTechs Data Recording Software requires no programming skills, and enables you to effortlessly select a reading rate, specify a user ID, and initiate the start of data collection. The data logger will be ready to use in just moments. In addition, all data can be saved in a format easily compatible with spreadsheet applications such as "Microsoft Excel" or "Lotus 1-2-3". It is our goal to bring you accurate, low-cost, easy-to-use data loggers that will easily integrate into your working environment. To better serve you and understand your needs, we welcome and appreciate your feedback. Thank you for choosing MadgeTech for your data logging requirements. Warranty Products manufactured by MadgeTech, Inc., are warranted against defective material and workmanship for a period of one year, starting from the date of shipment. In the event that a MadgeTech product is found to be defective, MadgeTech will repair or replace the product at its sole discretion. Such repair or replacement shall be the sole remedy of this warranty. This warranty extends only to the original purchasing customer and does not apply to any unit, which in our sole judgment, has been subjected to: Operating or environmental conditions in excess of our written specifications or recommendations; Damage, misuse or neglect; Improper installation, repair or alteration. This warranty excludes batteries and sensors. Except as to title, this is our only warranty for the products. MadgeTech, Inc. expressly disclaims all other warranties, guarantees or remedieswhether expressed or implied or statutoryincluding any implied warranty of merchantability or fitness for a particular purpose. We also disclaim any implied warranty arising out of trade usage or out of a course dealing or course of performance. We do not guarantee the integrity of data or warranty that the products will operate uninterrupted or error-free. Limitation of Liability MadgeTech data loggers and their associated software have been thoroughly tested and the documentation reviewed. However, MadgeTech does not warrant the performance of its products for any particular purpose. In no event is MadgeTech liable for any damages resulting, directly or indirectly, from the use of this or any products supplied by MadgeTech. Battery Warning See individual units for battery warning. 1 Quick Start To begin using your data logger right away, follow these six simple steps: Install the software. (See HYPERLINK \l "_Toc37140702"Software Installation) Attach the logger to your computer using the interface cable. (See HYPERLINK \l "_Toc37140704"Hardware Installation) From the Communication menu, select "Auto Configure Port". Fig 1.1 Fig 1.1  From the Device menu, select Start Device. Fig 1.3 Fig 1.3  A message will appear (fig 1.3) Click Yes to continue. Be sure to have downloaded all previously recorded data before continuing. Fig 1.3 The Start Device window will open. Note that there are two sections of this window, the Start Method section and the Start Parameters section. (Fig 1.4) Fig 1.4 The Start Method section allows the user to choose to Start Now or Delay Start, which allows most devices to record data anywhere from 0 to 60 days from the time it is started. The Start Parameters section allows the user to set three options in the device: User ID, Reading Rate, and Wrap Around. The User ID field allows the user to store a combination of six letters or numbers in the device to help you identify the device, or the person who started the device, etc. The Reading Rate field allows the user to select a specific reading rate in the device, typically ranging from one reading every 2 seconds to one reading every 12 hours. These ranges can vary depending on the device. The Wrap Around field is only available to devices which have the ability to perform memory wrap around. If the device is able to do so, the user chooses to overwrite the oldest data when the device memory becomes full. If this box is left unchecked, the device will stop recording data when the memory is full. When finished, click on Start Device. After a brief pause while the software communicates with the device, you will see the message "Device Started". Your device is now started and will begin taking measurements according to the parameters set in section 6. You may now disconnect the logger and place the device where you need it to perform its measurements. Later, when you are ready to view the measurements, simply reconnect it to the computer, and select "Read Device Data" from the Device menu. For more detailed information on using our software, please see the help file located under the help menu in the MadgeTech software. 2 Software Installation Installing from CD ROM Insert the CD ROM labeled MadgeTech Data Recording Software into your CD ROM drive. From the Windows Start menu, choose the Run command and type d:\setup into the Open field and click OK. If your CD ROM drive is not the D: drive, use the correct letter in the instructions above. After installation, the software will be listed under the default MadgeTech software program group and saved under the default C:\Program Files\MadgeTech 2.00\ directory. 3 Hardware Installation Connect the DB9 end of the interface cable provided with the software package to the computers serial port. (Fig 3.1 Fig. 3.1  Connect the stereo jack end of the interface cable to the data logger. Fig 3.2 Fig 3.2  CRYOTEMP To start the CRYOTEMP refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the CRYOTEMP is armed; data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. Once started, the LED will flash at the selected reading rate to indicate that the device is running. Be sure to keep the CRYOTEMP dry. If using the device in a dry ice environment, place the CRYOTEMP in a sealable plastic bag. If using the data logger in a wet ice environment, place the CRYOTEMP in a mason jar and close the cover. Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the screw in the back. Remove the battery from the battery sockets. If necessary, trim the new battery leads to match the length of the existing battery leads. Replace the battery with a Lithium LTC-7PN. Close case and secure with screw. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. OCTRTD To start the OCTRTD refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the OCTRTD is armed; data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. The OCTRTD has eight, four-position removable screw terminal connections, and can accept 2, 3 or 4-wire RTD configurations. Temperature probes to be used with the recorder must be 100 ohm platinum RTDs, in the standard 2-, 3- or 4-wire configuration. The OCTRTD is designed to achieve exceptional accuracy with the 4-wire probe, but will still yield measurements better than required for most applications with the 2- or 3-wire probes. Insure that the probe you select can be connected to the OCTRTD inputs by selecting probes with lead wires, or by attaching an adapter that will allow you to connect wire leads to the probe. OMEGAs PR-10 and PR-11 series of probes have the proper leads for direct connection to the OCTRTD. For 4-wire RTD probes, connect the four lead wires to the OCTRTD as shown in Fig. A. For 3-wire RTD probes, short inputs 3 and 4 together, then connect the lead wires to inputs 1, 2 and 3. For 2-wire RTD probes, short inputs 3 and 4 together and inputs 1 and 2 together, then connect the RTD lead wires to inputs 2 and 3. Fig. A  Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the four screws on the top of the case. Disconnect the battery from the battery connector. Replace the battery with a Lithium U9VL-J. Close the case and secure with screws, making sure that the wires for the battery connector do not get pinched between the cover and the case. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. OCTTEMP To start the OCTTEMP refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the OCTTEMP is armed; data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. The OCTTEMP has eight SMP connections which allow the user to insert subminiature thermocouple plugs into the connectors on the device. Fig. A shows how to connect the individual thermocouples. Fig. A  The software accepts eight different thermocouple types: J, K, T, E, R, S, B and N. To select the correct thermocouple type in the MadgeTech software: Select Thermocouple Type from the device menu. (Fig. B) Click on the change button in the Thermocouple Type window. Select the correct thermocouple type from the drop down list. Click on the Save button to store the thermocouple type in the device. Click Ok, to finish. Fig. B Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the four screws on the top of the case. Disconnect the battery from the battery connector. Replace the battery with a Lithium U9VL-J. Close the case and secure with screws, making sure that the wires for the battery connector do not get pinched between the cover and the case. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. QUADRTD To start the QUADRTD refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the QUADRTD is armed; data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. The QUADRTD has four, four-position removable screw terminal connections. The QUADRTD accepts 2, 3 or 4-wire RTD configurations. Temperature probe to be used with the recorder must be 100 ohm platinum RTDs, in the standard 2-, 3- or 4-wire configuration. The QUADRTD is designed to achieve exceptional accuracy with the 4-wire probe, but will still yield measurements better than required for most applications with the 2- or 3-wire probes. Insure that the probe you select can be connected to the QUADRTD input by selecting a probe with lead wires, or by attaching an adapter that will allow you to connect wire leads to the probe. OMEGAs PR-10 and PR-11 series of probes have the proper leads for direct connection to the QUADRTD. For 4-wire RTD probes, connect the four lead wires to the QUADRTD as shown in Fig. A. For 3-wire RTD probes, short inputs 3 and 4 together, then connect the lead wires to inputs 1, 2 and 3. For 2-wire RTD probes, short inputs 3 and 4 together and inputs 1 and 2 together, then connect the RTD lead wires to inputs 2 and 3. Fig. A  Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the four screws on the top of the case. Disconnect the battery from the battery connector. Replace the battery with a Lithium U9VL-J. Close the case and secure with screws, making sure that the wires for the battery connector do not get pinched between the cover and the case. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. QUADTEMP To start the QUADTEMP refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the QUADTEMP is armed; data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. The QUADTEMP has four SMP connections which allows the user to insert subminiature thermocouple plugs into the connectors on the device. The diagram below shows how to connect the individual thermocouples. Fig. A The software accepts eight different thermocouple types: J, K, T, E, R, S, B and N. To select the correct thermocouple type in the MadgeTech software: Select Thermocouple Type from the device menu. (Fig. B) Click on the change button in the Thermocouple Type window. Select the correct thermocouple type from the drop down list. Click on the Save button to store the thermocouple type in the device. Click Ok, to finish. Fig. B Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the four screws on the top of the case. Disconnect the battery from the battery connector. Replace the battery with a Lithium U9VL-J. Close the case and secure with screws, making sure that the wires for the battery connector do not get pinched between the cover and the case. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. RTDTEMP101 To start the RTDTEMP101 refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the RTDTEMP101 is armed; data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. The RTDTEMP101 has a four-position removable screw terminal connection. The RTDTEMP101 accepts 2, 3 or 4-wire RTD configurations. Temperature probe to be used with the recorder must be a 100 ohm platinum RTD, in the standard 2-, 3- or 4-wire configuration. The RTDTEMP101 is designed to achieve exceptional accuracy with the 4-wire probe, but will still yield measurements better than required for most applications with the 2- or 3-wire probes. Insure that the probe you select can be connected to the RTDTEMP101 input by selecting a probe with lead wires, or by attaching an adapter that will allow you to connect wire leads to the probe. OMEGAs PR-10 and PR-11 series of probes have the proper leads for direct connection to the RTDTEMP101. For 4-wire RTD probes, connect the four lead wires to the RTDTEMP101 as shown in Fig A. For 3-wire RTD probes, short inputs 3 and 4 together, then connect the lead wires to inputs 1, 2 and 3. For 2-wire RTD probes, short inputs 3 and 4 together and inputs 1 and 2 together, then connect the RTD lead wires to inputs 2 and 3. Fig A  Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the screw in the back. Remove the battery from the battery sockets. If necessary, trim the new battery leads to match the length of the existing battery leads. Replace the battery with a Lithium LTC-7PN. Close case and secure with screw. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. RTDTEMP110 To start the RTDTEMP110 refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the RTDTEMP110 is armed; data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. This device communicates at 57,600 baud. If you do not have the software set to Always Auto Configure, you must select this baud rate from the communication menu. The RTDTEMP110 has a four-position removable screw terminal connection. The RTDTEMP110 accepts 2, 3 or 4-wire RTD configurations. Temperature probe to be used with the recorder must be a 100 ohm platinum RTD, in the standard 2-, 3- or 4-wire configuration. The RTDTEMP110 is designed to achieve exceptional accuracy with the 4-wire probe, but will still yield measurements better than required for most applications with the 2- or 3-wire probes. Insure that the probe you select can be connected to the RTDTEMP110 input by selecting a probe with lead wires, or by attaching an adapter that will allow you to connect wire leads to the probe. OMEGAs PR-10 and PR-11 series of probes have the proper leads for direct connection to the RTDTEMP110. For 4-wire RTD probes, connect the four lead wires to the RTDTEMP110 as shown in Fig. A For 3-wire RTD probes, short inputs 3 and 4 together, then connect the lead wires to inputs 1, 2 and 3. For 2-wire RTD probes, short inputs 3 and 4 together and inputs 1 and 2 together, then connect the RTD lead wires to inputs 2 and 3. Fig. A  Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the screw in the back. Remove the battery from the battery sockets. If necessary, trim the new battery leads to match the length of the old battery leads. Replace the battery with a Lithium LTC-7PN. Close the case and secure with the screw. For 10+ year operation, be sure to set the reading rate to no more than one reading every 15 minutes. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. SAFETEMP To start the SAFETEMP without using the push button, refer to the section HYPERLINK \l "Quick_Start"Quick Start. Be sure that the Push Button Start button is not selected in the Start Device Window. To start the SAFETEMP using the Push Button, refer to HYPERLINK \l "Quick_Start"Quick Start and be sure that Push Button Start button is selected in the Start Device Window. After starting, the SAFETEMP is armed, but no data will be recorded. To start the device recording, the push button must be pressed and held down for 5 seconds or until both the Red and Green LEDs flash momentarily. The device will stop recording when the end of memory is reached or the push button has been held down for 5 seconds. At this point the device will only record data after it has been restarted. Once started, the green LED will flash every 10 seconds to indicate that the device is running. The red LED will flash if there is an alarm condition. To change the settings for the temperature alarm, select Alarm Settings from the Device menu in the MadgeTech software. A window will appear, allowing you to set a low and/or high temperature alarm as well as enable or disable an alarm. To clear an active alarm, press the clear alarm button. The SAFETEMP communicates at 600 baud. If you do not have the software set to Always Auto Configure, you must select this baud rate from the communication menu. The standard connector for the SAFETEMP is the SMP connection which allows for the user to insert the subminiature thermocouple plug into the connector on the device. The SAFETEMP will only accommodate a type K thermocouple. Fig A Fig. A Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the screw in the back. Remove the battery from the battery sockets. If necessary, trim the new battery leads to match the length of the old battery leads. Replace the battery with a Lithium LTC-7PN. Close the case and secure with the screw. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. TC110 To start the TC110 refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the TC110 is armed; data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. This device communicates at 57,600 baud. If you do not have the software set to Always Auto Configure, you must select this baud rate from the communication menu. The TC110 has a three position removable screw terminal. This enables the TC110 to be connected to most 2-wire thermocouples with lead wires, and to have the ability to shield the thermocouple with the ground connection. If the TC110 is to be used with a 2-wire thermocouple without shielding, leave the ground input unconnected. Be sure to connect the thermocouple with the right polarity as marked on the case of the device, the ground connection is optional. Fig A. Fig. A The software accepts eight different thermocouple types: J, K, T, E, R, S, B and N. To select the correct thermocouple type in the MadgeTech software to match the thermocouple that is being used: Select Thermocouple Type from the device menu. Fig. B Click on the change button in the Thermocouple Type window. Select the correct thermocouple type from the drop down list. Click on the Save button to store the thermocouple type in the device. Click Ok, to finish. Fig. B Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the screw in the back. Remove the battery from the battery sockets. If necessary, trim the new battery leads to match the length of the old battery leads. Replace the battery with a Lithium LTC-7PN. Close the case and secure with the screw. For 10+ year operation, be sure to set the reading rate to no more than one reading every 15 minutes. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. TC4000 To start the TC4000 refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the TC4000 is armed; data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. Depending on the physical configuration of the TC4000, the device will either have an input for a subminiature thermocouple connector or a removable screw terminal. The standard connector for the TC4000 is the SMP connection which allows for the user to insert a subminiature thermocouple plug into the connector on the device. Fig. A Fig. A The non-standard connector for the TC4000 is the removable screw terminal. This enables the TC4000 to be connected to most 2-wire thermocouples with lead wires. Be sure to connect the thermocouple with the right polarity as marked on the case of the device. Fig. B Fig. B The software accepts eight different thermocouple types: J, K, T, E, R, S, B and N. To select the correct thermocouple type in the MadgeTech software to match the thermocouple that is being used: Select Thermocouple Type from the device menu. Fig. C Click on the change button in the Thermocouple Type window. Select the correct thermocouple type from the drop down list. Click on the Save button to store the thermocouple type in the device. Click Ok, to finish. Fig. C Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the screw in the back. Remove the battery from the battery sockets. If necessary, trim the new battery leads to match the length of the old battery leads. Replace the battery with a Lithium LTC-7PN. Close the case and secure with the screw. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. TEMP100 To start the TEMP100 without using the push button, refer to the section HYPERLINK \l "Quick_Start"Quick Start. Be sure that the Push Button Start button is not selected in the Start Device Window. To start the TEMP100 using the Push Button, refer to HYPERLINK \l "Quick_Start"Quick Start and be sure that Push Button Start button is selected in the Start Device Window. After starting, the TEMP100 is armed, but no data will be recorded. To start the device recording, the push button must be pressed and held down for 5 seconds or until both the Red and Green LEDs flash momentarily. The device will stop taking data when the end of memory is reached or the push button has been held down for 5 seconds. At this point the device cannot be restarted until it has been re-armed by the computer. Once started, the green LED will flash every 10 seconds to indicate that the device is running. The red LED will flash if there is an alarm condition. To change the settings for the temperature alarm, select Alarm Settings from the Device menu in the MadgeTech software. A window will appear allowing you set a low and/or high temperature alarm as well as enable or disable the alarm. To clear an active alarm, press the clear alarm button. Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the screw in the back. Remove the battery from the battery sockets. If necessary, trim the new battery leads to match the length of the old battery leads. Replace the battery with a Lithium LTC-7PN. Close the case and secure with the screw. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. TEMP101 To start the TEMP101 refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the TEMP101 is armed; data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. Once started, the LED will flash at the selected reading rate to indicate that the device is running. The LED will flash in one second intervals if there is an alarm condition. To change the settings for the temperature alarm, select Alarm Settings from the Device menu in the MadgeTech software. A window will appear allowing you set a low and/or high temperature alarm as well as enable or disable the alarm. To clear an active alarm, press the clear alarm button. Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the screw in the back. Remove the battery from the battery sockets. If necessary, trim the new battery leads to match the length of the old battery leads. Replace the battery with a Lithium LTC-7PN. Close the case and secure with the screw. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. TEMP110 To start the TEMP110 refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the TEMP110 is armed, data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. Once started, the LED will flash at the selected reading rate to indicate that the device is running. This device communicates at 57,600 baud. If you do not have the software set to Always Auto Configure, you must select this baud rate from the communication menu. Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the screw in the back. Remove the battery from the battery sockets. If necessary, trim the new battery leads to match the length of the old battery leads. Replace the battery with a Lithium LTC-7PN. Close the case and secure with the screw. For 10+ year operation, be sure to set the reading rate to no more than one reading every 15 minutes. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. TEMP1000 The serial port on the TEMP1000 is located under the cap with the key ring. To connect the logger to the interface cable, unscrew the cap and insert the stereo jack end of the interface cable to the logger. To start the TEMP1000 refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the TEMP1000 is armed, data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Contact the factory on replacing the battery. Warning This device is intended for water proof applications. THIS UNIT MUST BE PROPERLY SEALED BEFORE SUBMERSING IN WATER OR OTHER MEDIUM. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. TEMP1000FP The serial port on the TEMP1000FP is located under the cap with the key ring. To connect the logger to the interface cable, unscrew the cap and insert the stereo jack end of the interface cable into the data logger. To start the TEMP1000FP refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the TEMP1000FP is armed, data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. Flexible Probe The maximum bending radius of the probe is 2, bending only once. Avoid bending the probe anywhere along the first 2.5 from the tip. Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Contact the factory for information on replacing the battery. Warning This device is intended for water proof applications. THIS UNIT MUST BE PROPERLY SEALED BEFORE SUBMERSING IN WATER OR OTHER MEDIUM. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. TEMP1000IS The serial port on the TEMP1000IS is located under the cap with the key ring. To connect the logger to the interface cable, unscrew the cap and insert the stereo jack end of the interface cable into the logger. To start the TEMP1000IS refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the TEMP1000IS is armed, data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Contact the factory for information on replacing the battery. Warning This device is intended for water proof applications. THIS UNIT MUST BE PROPERLY SEALED BEFORE SUBMERSING IN WATER OR OTHER MEDIUM. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. TEMP1000P The serial port on the TEMP1000P is located under the cap with the key ring. To connect the logger to the interface cable, unscrew the cap and insert the stereo jack end of the interface cable into the logger. To start the TEMP1000P refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the TEMP1000P is armed, data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Contact the factory for information on replacing the battery. Warning This device is intended for water proof applications. THIS UNIT MUST BE PROPERLY SEALED BEFORE SUBMERSING IN WATER OR OTHER MEDIUM. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. TEMP1000S The serial port on the TEMP1000S is located under the cap with the key ring. To connect the logger to the interface cable, unscrew the cap and insert the stereo jack end of the interface cable into the logger. To start the TEMP1000S refer to  HYPERLINK \l "Quick_Start" Quick Start. After starting, the TEMP1000S is armed, data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Contact the factory for information on replacing the battery. Warning This device is intended for water proof applications. THIS UNIT MUST BE PROPERLY SEALED BEFORE SUBMERSING IN WATER OR OTHER MEDIUM. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. TEMPRETRIEVER To start the TEMPRETRIEVER refer to the section  HYPERLINK \l "_Toc37140701" HYPERLINK \l "Quick_Start"Quick Start. After starting, the TEMP101 is armed; data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. Once started, the LED will flash at the selected reading rate to indicate that the device is running. The LED will flash in one second intervals if there is an alarm condition. To change the settings for the temperature alarm, select Alarm Settings from the Device menu in the MadgeTech software. A window will appear allowing you set a low and/or high temperature alarm as well as enable or disable the alarm. To clear an active alarm, press the clear alarm button. Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the screw in the back. Remove the battery from the battery sockets. If necessary, trim the new battery leads to match the length of the old battery leads. Replace the battery with a Lithium LTC-7PN. Close the case and secure with the screw. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. THERMOVAULT The THERMOVAULT consists of a TC4000 with a fixed screw terminal connector, and a thermally insulated box (logger box) to house the TC4000. In order to communicate with the THERMOVAULT, the TC4000 must be removed from the box in order to connect the IFC101 to the TC4000. Follow these instructions on using the THERMOVAULT: To start the TC4000 refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the TC4000 is armed; data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. Place the data logger into the box. Wind the thermocouple around the small channel in the logger box. The end of the thermocouple should extend out the edge of the logger box. Place the logger box in oven or on oven conveyor. The oven temperature determines the maximum time the logger box can remain in the oven. The following table shows the maximum duration the box may remain in the oven. Ambient Temperature, C (F)Max Duration in Minutes100 (212)52150 (302)30200 (392)22250 (482)17260 (500)16300 (572)15350 (662)12400 (752)10 Leaving the logger box in the oven longer than stated above or immersing logger box into any liquid, could cause loss of data, damage to data logger or the logger box. Remove data logger from the logger box immediately after removing from oven. The outside of the case will be hot, protective gloves or tongs must be used to handle the box. Failure to remove the data logger from the logger box could cause loss of data or damage to the data logger. Allow the logger box to cool to room temperature before using again. Failure to cool the box to room temperature before using again could cause loss to data, damage to the data logger or damage to the logger box. Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Remove the TC4000 from the logger box. Open the case by removing the screw in the back. Remove the battery from the battery sockets. If necessary, trim the new battery leads to match the length of the existing battery leads. Replace the battery with a Lithium LTC-7PN. Close the case and secure with the screw. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. MICRORHTEMP To start the MICRORHTEMP, refer to the section HYPERLINK \l "Quick_Start"Quick Start. After starting, the MICRORHTEMP is armed; data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. Once started, the green LED will flash at the selected reading rate to indicate that the device is running. The red LED will flash in one second intervals if there is an alarm condition. To change the settings for the temperature alarm, select Alarm Settings from the Device menu in the MadgeTech software. A window will appear allowing you set a low and/or high temperature alarm as well as enable or disable the alarm. To clear an active alarm, press the clear alarm button. Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Unscrew the cap at the top of the case. Remove the two coin cell batteries; be sure not to remove the device from the case. Insert two Lithium CR1225 batteries by placing the positive ends (flat surface) away from the device, leave enough slack in the black wire to let the spring make contact with the positive end of the battery. Close the case by screwing on the cap. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. MICRORHTEMP-EB To start the MICRORHTEMP-EB, refer to the section HYPERLINK \l "Quick_Start"Quick Start. After starting, the MICRORHTEMP-EB is armed; data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. Once started, the green LED will flash at the selected reading rate to indicate that the device is running. The red LED will flash in one second intervals if there is an alarm condition. To change the settings for the temperature alarm, select Alarm Settings from the Device menu in the MadgeTech software. A window will appear allowing you set a low and/or high temperature alarm as well as enable or disable the alarm. To clear an active alarm, press the clear alarm button. Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Unscrew the cap at the top of the case. Remove the two coin cell batteries; be sure not to remove the device from the case. Insert two Lithium CR1632 batteries by placing the positive ends (flat surface) away from the device, leave enough slack in the black wire to let the spring make contact with the positive end of the battery. Close the case by screwing on the cap. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. RHTEMP101 To start the RHTEMP101 refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the RHTEMP101 is armed; data will start to recording. The device will stop recording when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. Once started, the LED will flash at the selected reading rate to indicate that the device is running. The LED will flash in one second intervals if there is an alarm condition. To change the settings for the temperature alarm, select Alarm Settings from the Device menu in the MadgeTech software. A window will appear allowing you set a low and/or high temperature alarm as well as enable or disable the alarm. To clear an active alarm, press the clear alarm button. Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the screw in the back. Remove the battery from the battery sockets. If necessary, trim the new battery leads to match the length of the existing battery leads. Replace the battery with a Lithium LTC-3PN. Close the case and secure with the screw. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. RHTEMP110 To start the RHTEMP110 refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the RHTEMP110 is armed, data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. Once started, the LED will flash at the selected reading rate to indicate that the device is running. This device communicates at 57,600 baud. If you do not have the software set to Always Auto Configure, you must select this baud rate from the communication menu. Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the screw in the back. Remove the battery from the battery sockets. If necessary, trim the new battery leads to match the length of the existing battery leads. Replace the battery with a Lithium LTC-7PN. Close the case and secure with the screw. For 10+ year operation, be sure to set the reading rate to no more than one reading every 15 minutes. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. RHTEMP1000 The serial port on the RHTEMP1000 is located under the cap with the key ring. To connect the logger to the interface cable, unscrew the cap and insert the stereo jack end of the interface cable into the logger. To start the RHTEMP1000 refer to HYPERLINK \l "Quick_Start"Quick Start. After starting the RHTEMP1000, it is armed and data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. For humidity readings, unscrew the cap enough to expose the humidity sensor to the air. The cap is fully open when it becomes difficult to turn. Do not allow water or other liquids to come into contact with the inside of the sensor compartment. If the device needs to be water tight, screw the cap on as tightly as possible.  Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Contact the factory for information on replacing the battery. Warning This device is intended for water proof applications. THIS UNIT MUST BE PROPERLY SEALED BEFORE SUBMERSING IN WATER OR OTHER MEDIUM. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. RHTEMP1000IS The serial port on the RHTEMP1000IS is located under the cap with the key ring. To connect the logger to the interface cable, unscrew the cap and insert the stereo jack end of the interface cable into the logger. To start the RHTEMP1000IS refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the RHTEMP1000IS is armed, data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. For humidity readings, unscrew the cap enough to expose the humidity sensor to the air. The cap is fully open when it becomes difficult to turn. Do not allow water or other liquids to come into contact with the inside of the sensor compartment. If the device needs to be water tight, screw the cap on as tightly as possible.  Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Contact the factory for information on replacing the battery. Warning This device is intended for water proof applications. THIS UNIT MUST BE PROPERLY SEALED BEFORE SUBMERSING IN WATER OR OTHER MEDIUM. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. OCTVOLT To start the OCTVOLT refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the OCTVOLT is armed; data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. Once started, the LED will flash at the selected reading rate to indicate that the device is running. The OCTVOLT allows for user defined units to be programmed into the device. To use the programmable engineering units, refer to the  HYPERLINK \l "Engineering_Units" Engineering Units section. The OCTVOLT has eight, two-position removable screw terminal connections. The OCTVOLT accepts 2-wire configurations. Insure that the device that is to be measured can be connected to the OCTVOLT input by selecting an output connection with lead wires, or by attaching an adapter that will allow you to connect wire leads to the device to be measured. For a two-wire connection, connect IN+ to positive signal (typically the red wire) and IN- to the negative signal (typically the black wire). Fig. A Fig. A  Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the four screws on the top of the case. Disconnect the battery from the battery connector. Replace the battery with a Lithium U9VL-J. Close the case and secure with screws, making sure that the wires for the battery connector do not get pinched between the cover and the case. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. QUADVOLT To start the QUADVOLT refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the QUADVOLT is armed; data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. Once started, the LED will flash at the selected reading rate to indicate that the device is running. The QUADVOLT allows for user defined units to be programmed into the device. To use the programmable engineering units, refer to the  HYPERLINK \l "Engineering_Units" Engineering Units section. The QUADVOLT has four two-position removable screw terminal connections. The QUADVOLT accepts 2-wire configurations. Insure that the device that is to be measured can be connected to the QUADVOLT input by selecting an output connection with lead wires, or by attaching an adapter that will allow you to connect wire leads to the device to be measured. For a two-wire connection, connect IN+ to positive signal (typically the red wire) and IN- to the negative signal (typically the black wire). Fig. A Fig. A Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the four screws on the top of the case. Disconnect the battery from the battery connector. Replace the battery with a Lithium U9VL-J. Close the case and secure with screws, making sure that the wires for the battery connector do not get pinched between the cover and the case. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. VOLT101 To start the VOLT101 refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the VOLT101 is armed; data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. Once started, the LED will flash at the selected reading rate to indicate that the device is running. The VOLT101 allows for user defined units to be programmed into the device. To use the programmable engineering units, refer to the  HYPERLINK \l "Engineering_Units" Engineering Units section. (page ?) The VOLT101 has a two-position removable screw terminal connection. The VOLT101 accepts 2-wire configurations. Insure that the device that is to be measured can be connected to the VOLT101 input by selecting an output connection with lead wires, or by attaching an adapter that will allow you to connect wire leads to the device to be measured. For a two-wire connection, connect IN+ to positive signal (typically the red wire) and IN- to the negative signal (typically the black wire). Fig. A Fig. A Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the screw in the back. Remove the battery from the battery sockets. If necessary, trim the new battery leads to match the length of the existing battery leads. Replace the battery with a Lithium LTC-7PN. Close the case and secure with the screw. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. VOLT110 100mV To start the VOLT110 100 mV refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the VOLT110 100 mV is armed; data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. This device communicates at 57,600 baud. If you do not have the software set to Always Auto Configure, you must select this baud rate from the communication menu. Once started, the LED will flash at the selected reading rate to indicate that the device is running. The VOLT110 100 mV allows for user defined units to be programmed into the device. To use the programmable engineering units, refer to the  HYPERLINK \l "Engineering_Units" Engineering Units section. (page ?) The VOLT110 100 mV has a three-position removable screw terminal connection. The VOLT110 100 mV measures a differential input and accepts 2 or 3-wire configurations. Insure that the device that is to be measured can be connected to the VOLT110 100 mV input by selecting an output connection with lead wires, or by attaching an adapter that will allow you to connect wire leads to the device to be measured. For a two-wire connection, connect IN+ to positive signal (typically the red wire) and IN- to the negative signal (typically the black wire). For a three-wire connection, make the connections as described for the two-wire connection, also connecting Ground to signal ground. Fig. A Fig. A  Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the screw in the back. Remove the battery from the battery sockets. If necessary, trim the new battery leads to match the length of the existing battery leads. Replace the battery with a Lithium LTC-7PN. Close the case and secure with the screw. For 10+ year operation, be sure to set the reading rate to no more than one reading every 15 minutes. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. VOLT110 2.5, 15 and 30 V To start the VOLT110 refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the VOLT110 is armed; data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. This device communicates at 57,600 baud. If you do not have the software set to Always Auto Configure, you must select this baud rate from the communication menu. Once started, the LED will flash at the selected reading rate to indicate that the device is running. The VOLT110 allows for user defined units to be programmed into the device. To use the programmable engineering units, refer to the  HYPERLINK \l "Engineering_Units" Engineering Units section. The VOLT110 has a two-position removable screw terminal connection. The VOLT110 accepts 2-wire configurations. Insure that the device that is to be measured can be connected to the VOLT110 input by selecting an output connection with lead wires, or by attaching an adapter that will allow you to connect wire leads to the device to be measured. For a two-wire connection, connect IN+ to positive signal (typically the red wire) and IN- to the negative signal (typically the black wire). Fig. A Fig. A Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the screw in the back. Remove the battery from the battery sockets. If necessary, trim the new battery leads to match the length of the existing battery leads. Replace the battery with a Lithium LTC-7PN. Close the case and secure with the screw. For 10+ year operation, be sure to set the reading rate to no more than one reading every 15 minutes. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. OCTPROCESS To start the OCTPROCESS refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the OCTPROCESS is armed; data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. The OCTPROCESS allows for user defined units to be programmed into the device. To use the programmable engineering units, refer to the  HYPERLINK \l "Engineering_Units" Engineering Units section. (page ?) The OCTPROCESS has eight two-position removable screw terminal connections. The OCTPROCESS accepts 2-wire configurations. Insure that the device that is to be measured can be connected to the OCTPROCESS input by selecting an output connection with lead wires, or by attaching an adapter that will allow you to connect wire leads to the device to be measured. For a two-wire loop, connect IN+ to positive signal (typically the red wire) and IN- to the negative signal (typically the black wire). Fig. A Fig. A Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the four screws on the top of the case. Disconnect the battery from the battery connector. Replace the battery with a Lithium U9VL-J. Close the case and secure with screws, making sure that the wires for the battery connector do not get pinched between the cover and the case. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. PROCESS101 To start the PROCESS101 refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the PROCESS101 is armed; data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. Once started, the LED will flash at the selected reading rate to indicate that the device is running. The PROCESS101 allows for user defined units to be programmed into the device. To use the programmable engineering units, refer to the  HYPERLINK \l "Engineering_Units" Engineering Units section. The PROCESS101 has a two-position removable screw terminal connection. The PROCESS101 accepts 2-wire configurations. Insure that the device that is to be measured can be connected to the PROCESS101 input by selecting an output connection with lead wires, or by attaching an adapter that will allow you to connect wire leads to the device to be measured. For a two-wire loop, connect IN+ to positive signal (typically the red wire) and IN- to the negative signal (typically the black wire). Fig. A Fig A Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the screw in the back. Remove the battery from the battery sockets. If necessary, trim the new battery leads to match the length of the existing battery leads. Replace the battery with a Lithium LTC-7PN. Close the case and secure with the screw. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. PROCESS110 To start the PROCESS110 refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the PROCESS110 is armed; data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. This device communicates at 57,600 baud. If you do not have the software set to Always Auto Configure, you must select this baud rate from the communication menu. Once started, the LED will flash at the selected reading rate to indicate that the device is running. The PROCESS110 allows for user defined units to be programmed into the device. To use the programmable engineering units, refer to the  HYPERLINK \l "Engineering_Units" Engineering Units section. The PROCESS110 has a three-position removable screw terminal connection. The PROCESS110 accepts 2 or 3-wire configurations. Insure that the device that is to be measured can be connected to the PROCESS110 input by selecting an output connection with lead wires, or by attaching an adapter that will allow you to connect wire leads to the device to be measured. For a two-wire loop, connect IN+ to positive signal (typically the red wire) and IN- to the negative signal (typically the black wire). You can leave ground unconnected, or connect it to signal shield. Fig. A Fig. A  Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the screw in the back. Remove the battery from the battery sockets. If necessary, trim the new battery leads to match the length of the existing battery leads. Replace the battery with a Lithium LTC-7PN. Close the case and secure with the screw. For 10+ year operation, be sure to set the reading rate to no more than one reading every 15 minutes. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. QUADPROCESS To start the QUADPROCESS refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the QUADPROCESS is armed; data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. The QUADPROCESS allows for user defined units to be programmed into the device. To use the programmable engineering units, refer to the  HYPERLINK \l "Engineering_Units" Engineering Units section. (page ?) The QUADPROCESS has four, two-position removable screw terminal connections. The QUADPROCESS accepts 2-wire configurations. Insure that the device that is to be measured can be connected to the QUADPROCESS input by selecting an output connection with lead wires, or by attaching an adapter that will allow you to connect wire leads to the device to be measured For a two-wire loop, connect IN+ to positive signal (typically the red wire) and IN- to the negative signal (typically the black wire). Fig. A Fig. A  Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the four screws on the top of the case. Disconnect the battery from the battery connector. Replace the battery with a Lithium U9VL-J. Close the case and secure with screws, making sure that the wires for the battery connector do not get pinched between the cover and the case. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. EVENT101 To start the EVENT101 refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the EVENT101 is armed; data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. Once started, the LED will flash at the selected reading rate to indicate that the device is running. The EVENT101 has a two-position removable screw terminal connection. The EVENT101 accepts 2-wire configurations. To use, connect IN+ to positive signal (typically the red wire) and IN- to the negative signal (typically the black wire). Fig. A Fig. A The EVENT101 records the time at which a pulse input or contact closure occurs on the input of the device. The device is continuously sampling at 1 kHz, therefore, the pulse input must be at least 1 ms in duration. An input voltage of greater than 30 volts on the input could cause loss of data or damage to data logger. Only one event will be recorded per reading interval (ie. 1 second). The EVENT101 will most often be used with the 1 second reading rate. Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the screw in the back. Remove the battery from the battery sockets. If necessary, trim the new battery leads to match the length of the existing battery leads. Replace the battery with a Lithium LTC-7PN. Close the case and secure with the screw. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. EVENT110 To start the EVENT110 refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the EVENT110 is armed; data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. Once started, the LED will flash at the selected reading rate to indicate that the device is running. This device communicates at 57,600 baud. If you do not have the software set to Always Auto Configure, you must select this baud rate from the communication menu. The EVENT110 has a two-position removable screw terminal connection. The EVENT110 accepts 2-wire configurations. To use, connect IN+ to positive signal (typically the red wire) and IN- to the negative signal (typically the black wire). Fig A Fig. A The EVENT110 records the time at which a pulse input or contact closure occurs on the input of the device. The device is continuously sampling at 1 kHz, therefore, the pulse input must be at least 1 ms in duration. An input voltage of greater than 30 volts on the input could cause loss of data or damage to data logger. Only one event will be recorded per reading interval (i.e. 1 second). The EVENT110 will most often be used with the 1 second reading rate. Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the screw in the back. Remove the battery from the battery sockets. If necessary, trim the new battery leads to match the length of the existing battery leads. Replace the battery with a Lithium LTC-7PN. Close the case and secure with the screw. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. PULSE101 To start the PULSE101 refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the PULSE101 is armed; data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. Once started, the LED will flash at the selected reading rate to indicate that the device is running. The PULSE101 allows for user defined units to be programmed into the device. To use the programmable engineering units, refer to the Engineering Units section. (page ?) The PULSE101 has a two-position removable screw terminal connection. The PULSE101 accepts 2-wire configurations. To use, connect IN+ to positive signal (typically the red wire) and IN- to the negative signal (typically the black wire). Fig. A Fig. A The PULSE101 records the number of pulse inputs or contact closures on the input of the device. The input is continuously sampled at 1 kHz, therefore, the input pulse must be at least 1 ms in duration. This device will read up to 100 pulses per second. The counter is reset at the beginning, and is recorded at the end of each sample interval. The counter is incremented when the input transitions high during the reading interval. Note: An input voltage of greater than 30 volts on the input could cause loss of data or damage to data logger. Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the screw in the back. Remove the battery from the battery sockets. If necessary, trim the new battery leads to match the length of the existing battery leads. Replace the battery with a Lithium LTC-7PN. Close the case and secure with the screw. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. PULSE110 To start the PULSE110 refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the PULSE110 is armed; data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. Once started, the LED will flash at the selected reading rate to indicate that the device is running. This device communicates at 57,600 baud. If you do not have the software set to Always Auto Configure, you must select this baud rate from the communication menu. The PULSE110 allows for user defined units to be programmed into the device. To use the programmable engineering units, refer to the Engineering Units section. (page ?) The PULSE110 has a two-position removable screw terminal connection. The PULSE110 accepts 2-wire configurations. To use, connect IN+ to positive signal (typically the red wire) and IN- to the negative signal (typically the black wire). Fig. A Fig. A The PULSE110 records the number of pulse inputs or contact closures on the input of the device. The input is continuously sampled at 1 kHz, therefore, the input pulse must be at least 1 ms in duration. This device will read up to 100 pulses per second. The counter is reset at the beginning, and is recorded at the end of each sample interval. The counter is incremented when the input transitions high during the reading interval. Note: An input voltage of greater than 30 volts on the input could cause loss of data or damage to data logger. Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the screw in the back. Remove the battery from the battery sockets. If necessary, trim the new battery leads to match the length of the existing battery leads. Replace the battery with a Lithium LTC-7PN. Close the case and secure with the screw. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. STATE101 To start the STATE101 refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the STATE101 is armed; data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. Once started, the LED will flash at the selected reading rate to indicate that the device is running. The STATE101 has a two-position removable screw terminal connection. The STATE101 accepts 2-wire configurations. To use, connect IN+ to positive signal (typically the red wire) and IN- to the negative signal (typically the black wire). Fig. A Fig. A The STATE101 records the state of the input of the device at the selected reading rate. To ensure that the STATE101 captures the state of the input, be sure to have a contact closure that lasts at least the length of the selected reading rate. The device will only record if the state has changed when the input is sampled. The date and time is recorded along with the change of state. Note: An input voltage of greater than 30 volts on the input could cause loss of data or damage to data logger. Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the screw in the back. Remove the battery from the battery sockets. If necessary, trim the new battery leads to match the length of the existing battery leads. Replace the battery with a Lithium LTC-7PN. Close the case and secure with the screw. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. STATE110 To start the STATE110 refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the STATE110 is armed; data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. Once started, the LED will flash at the selected reading rate to indicate that the device is running. This device communicates at 57,600 baud. If you do not have the software set to Always Auto Configure, you must select this baud rate from the communication menu. The STATE110 has a two-position removable screw terminal connection. The STATE110 accepts 2-wire configurations. To use, connect IN+ to positive signal (typically the red wire) and IN- to the negative signal (typically the black wire). Fig. A Fig. A The STATE110 records the state of the input of the device at the selected reading rate. To ensure that the STATE110 captures the state of the input, be sure to have a contact closure that lasts at least the length of the selected reading rate. The device will only record if the state has changed when the input is sampled. The date and time is recorded along with the change of state. Note: An input voltage of greater than 30 volts on the input could cause loss of data or damage to data logger. Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the screw in the back. Remove the battery from the battery sockets. If necessary, trim the new battery leads to match the length of the existing battery leads. Replace the battery with a Lithium LTC-7PN. Close the case and secure with the screw. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. PRHTEMP101 To start the PRHTEMP101 refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the PRHTEMP101 is armed; data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. Once started, the LED will flash at the selected reading rate to indicate that the device is running. Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the screw in the back. Remove the battery from the battery sockets. If necessary, trim the new battery leads to match the length of the existing battery leads. Replace the battery with a Lithium LTC-7PN. Close the case and secure with the screw. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. PRHTEMP110 To start the PRHTEMP110 refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the PRHTEMP110 is armed, data will start recording. The device will stop recording when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. Once started, the LED will flash at the selected reading rate to indicate that the device is running. This device communicates at 57,600 baud. If you do not have the software set to Always Auto Configure, you must select this baud rate from the communication menu. Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the screw in the back. Remove the battery from the battery sockets. If necessary, trim the new battery leads to match the length of the existing battery leads. Replace the battery with a Lithium LTC-7PN. Close the case and secure with the screw. For 10+ year operation, be sure to set the reading rate to no more than one reading every 15 minutes. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. PRTEMP101 To start the PRTEMP101 refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the PRTEMP101 is armed; data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. Once started, the LED will flash at the selected reading rate to indicate that the device is running. Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the screw in the back. Remove the battery from the battery sockets. If necessary, trim the new battery leads to match the length of the existing battery leads. Replace the battery with a Lithium LTC-7PN. Close the case and secure with the screw. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. PRTEMP110 To start the PRTEMP110 refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the PRTEMP110 is armed, data will start recording. The device will stop recording when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. Once started, the LED will flash at the selected reading rate to indicate that the device is running. This device communicates at 57,600 baud. If you do not have the software set to Always Auto Configure, you must select this baud rate from the communication menu. Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the screw in the back. Remove the battery from the battery sockets. If necessary, trim the new battery leads to match the length of the existing battery leads. Replace the battery with a Lithium LTC-7PN. Close the case and secure with the screw. For 10+ year operation, be sure to set the reading rate to no more than one reading every 15 minutes. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. PRTEMP1000 The serial port on the PRTEMP1000 is located under the cap with the key ring. To connect the logger to the interface cable, unscrew the cap and insert the stereo jack end of the interface cable into the logger. To start the PRTEMP1000 refer to HYPERLINK \l "Quick_Start"Quick Start. After starting the PRTEMP1000, it is armed and data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. To use the pressure sensor for gauge measurements, screw the male NPT fitting into the pipe to be measured. The NPT fitting can also be used for absolute and submersible applications, but it is preferable to use the submersible fitting for these applications. With NPT fitting With submersible fitting  Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Contact the factory for information on replacing the battery. Warning This device is intended for water proof applications. THIS UNIT MUST BE PROPERLY SEALED BEFORE SUBMERSING IN WATER OR OTHER MEDIUM. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. PRTEMP1000P The serial port on the PRTEMP1000 is located under the cap with the key ring. To connect the logger to the interface cable, unscrew the cap and insert the stereo jack end of the interface cable into the logger. To start the PRTEMP1000 refer to HYPERLINK \l "Quick_Start"Quick Start. After starting the PRTEMP1000, it is armed and data will start recording. The device will stop recording when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Contact the factory for information on replacing the battery. Warning This device is intended for water proof applications. THIS UNIT MUST BE PROPERLY SEALED BEFORE SUBMERSING IN WATER OR OTHER MEDIUM. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. pHTEMP101 To start the pHTEMP101 refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the pHTEMP101 is armed, data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. Although the pHTEMP101 is used in essentially the same manner as the rest of the MadgeTech data loggers, a few extra considerations are necessary to achieve the best performance. Determination of pH requires the measurement of two parameters: electrical potential and temperature. Thus, the pHTemp101 has inputs for two external probes: a BNC jack that will accept most standard pH electrodes, and a 4-position header block that can be easily configured to accept a variety of temperature probes. To use the pHTEMP101 data recorder: The selected pH electrode to be used with the recorder should have a BNC output connection, or an appropriate adapter. To minimize measurement error, you should select a probe with an output impedance less than 300 megaohms at the desired temperature. The OMEGA PHE-4200 Series of probes satisfy this condition above 0 C (32 F), and have a BNC output connection. The selected temperature probe to be used with the recorder must be a 100 ohm platinum RTD, in the standard 2-,3- or 4-wire configuration. The pHTEMP101 is designed to achieve exceptional accuracy with the 4-wire probe, but will still yield measurements better than required for a pH-measurement with the 2- or 3-wire probes. Insure that the probe you select can be connected to the Phtemp101 RTD input by selecting a probe with lead wires, or by attaching an adapter that will allow you to connect wire leads to the probe. OMEGAs PR-10 and PR-11 series of probes have the proper leads for direct connection to the Phtemp101. Connect the probes to the data logger. Fig. A Refer to the description of your pH probe for an applicable calibration procedure. You may now use the pHTEMP101 exactly as you would any other MadgeTech data loggers. Fig. A  Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the screw in the back. Remove the battery from the battery sockets. Replace the battery with a Lithium U9VL-J. Close the case and secure with the screw. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. LEVEL1000 The serial port on the LEVEL1000 is located under the cap with the key ring. To connect the logger to the interface cable, unscrew the cap and insert the stereo jack end of the interface cable into the logger. To start the LEVEL1000 refer to HYPERLINK \l "Quick_Start"Quick Start. After starting the LEVEL1000, it is armed and data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. This device is not atmospheric pressure compensated, therefore, a reading should be taken prior to submerging the device into water. That reading will then need to be subtracted from all readings to provide accurate water level measurement. When submersing into water, be sure to connect the key ring at the top of the logger to a secure line. Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Contact the factory for information on replacing the battery. Warning This device is intended for water proof applications. THIS UNIT MUST BE PROPERLY SEALED BEFORE SUBMERSING IN WATER OR OTHER MEDIUM. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. LEVEL2000 The serial port on the LEVEL2000 is located under the cap with the key ring. To connect the logger to the interface cable, unscrew the cap and insert the stereo jack end of the interface cable into the logger. To start the LEVEL2000 refer to HYPERLINK \l "Quick_Start"Quick Start. After starting the LEVEL2000, it is armed and data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. This device is atmospheric pressure compensated; therefore, the end of the tubing must remain above water. The tubing also contains a serial cable for connecting the data logger to a pc; do not allow water or other liquids to come into contact with the end of the serial cable. When submersing into water, be sure to connect the key ring at the top of the logger to a secure line. Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Contact the factory for information on replacing the battery. Warning This device is intended for water proof applications. THIS UNIT MUST BE PROPERLY SEALED BEFORE SUBMERSING IN WATER OR OTHER MEDIUM. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. SHOCK101 To start the SHOCK101 refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the SHOCK101 is armed, data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. The SHOCK101 records peak acceleration levels for a specified reading interval. This means that the SHOCK101 will record the largest acceleration level on each of the axiss during each reading interval. For example, if the reading rate is set at 2 seconds, then the largest acceleration value during the 2 seconds will be recorded for each of the x, y and z axiss. Due to the noise of the electronics, there is typically 0.2 to 0.3 g of noise on each of the axiss. During real-time operation, the SHOCK101 will only take one measurement during the reading interval, therefore in real-time operation, the device is not measuring the peak acceleration levels. When setting up the SHOCK101, note which way the axiss are to determine which direction the shock or acceleration occurs. When the SHOCK101 is laying flat with the label facing up, the x-axis is the one that reads left to right and will read around 0 when still. The y-axis reads top to bottom and will read around 0 when still. The z-axis reads perpendicular to the other two axiss, and will read around 1 when still. Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the four screws on the top of the case. Disconnect the battery from the battery connector. Replace the battery with a Lithium U9VL-J. Close the case and secure with screws, making sure that the wires for the battery connector do not get pinched between the cover and the case. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. SHOCK101-EB To start the SHOCK101-EB, refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the SHOCK101-EB is armed, data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. The SHOCK101-EB records peak acceleration levels for a specified reading interval. This means that the SHOCK101-EB will record the largest acceleration level on each of the axiss during each reading interval. For example, if the reading rate is set at 2 seconds, then the largest acceleration value during the 2 seconds will be recorded for each of the x, y and z axiss. Due to the noise of the electronics, there is typically 0.2 to 0.3 g of noise on each of the axiss. During real-time operation, the SHOCK101-EB will only take one measurement during the reading interval, therefore in real-time operation, the device is not measuring the peak acceleration levels. When setting up the SHOCK101-EB, notice which way the axiss are to determine which direction the shock or acceleration occurs. When the SHOCK101-EB is laying flat with the label facing up, the x-axis is the one that reads left to right and will read around 0 when still. The y-axis reads top to bottom and will read around 0 when still. The z-axis reads perpendicular to the other two axiss, and will read around 1 when still. Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the four screws on the bottom of the large case. Disconnect the battery connector to allow the cover to completely open. Remove the 6 D-cell batteries from the holder. Replace the batteries with 6 D-cell batteries. Reconnect the battery connector, close the case and secure with the screws. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. ULTRASHOCK To start the ULTRASHOCK refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the ULTRASHOCK is armed, data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. The ULTRASHOCK records peak acceleration levels for a specified reading interval. This means that the ULTRASHOCK will record the largest acceleration level on each of the axiss during each reading interval. For example, if the reading rate is set at 2 seconds, then the largest acceleration value during the 2 seconds will be recorded for each of the x, y and z axiss. Due to the noise of the electronics, there is typically 0.2 to 0.3 g of noise on each of the axiss. During real-time operation, the ULTRASHOCK will only take one measurement during the reading interval, therefore in real-time operation, the device not measuring the peak acceleration levels. When setting up the ULTRASHOCK, notice which way the axiss are to determine which direction the shock or acceleration occurs. When the ULTRASHOCK is laying flat with the label facing up, the x-axis is the one that reads left to right and will read around 0 when still. The y-axis reads top to bottom and will read around 0 when still. The z-axis reads perpendicular to the other two axiss, and will read around 1 when still. Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the four screws on the top of the case. Disconnect the battery from the battery connector. Replace the battery with a Lithium U9VL-J. Close the case and secure with screws, making sure that the wires for the battery connector do not get pinched between the cover and the case. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. PHONEPROCESS The PHONEPROCESS differs from other MadgeTech data loggers by communicating through an onboard modem. In order to communicate to the PHONEPROCESS, the data logger must be connected to a phone line. To use the software to connect to the data logger, the correct COM port must be selected in our software. If you are unsure which COM port the modem is on, you can use Windows Device Manager to locate the port number. The PHONEPROCESS communicates at 2400 baud; therefore, the correct baud rate in the software must be selected. To connect, use the Modem Connection feature of the MadgeTech software located under the communication menu. In the telephone number field, enter the telephone number of the line that the PHONEPROCESS is on. The PHONEPROCESS will then answer the call and connect to the MadgeTech software. To start the PHONEPROCESS refer to HYPERLINK \l "Quick_Start"Quick Start and skip ahead to section 4. After starting, the PHONEPROCESS is armed; data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. Once started, the LED will flash at the selected reading rate to indicate that the device is running. The PHONEPROCESS allows for user defined units to be programmed into the device. To use the programmable engineering units, refer to the  HYPERLINK \l "Engineering_Units" Engineering Units section. (page ?) The PHONEPROCESS has a two-position removable screw terminal connection. The PHONEPROCESS accepts 2-wire configurations. Insure that the device that is to be measured can be connected to the PHONEPROCESS input by selecting an output connection with lead wires, or by attaching an adapter that will allow you to connect wire leads to the device to be measured. For a two-wire loop, connect IN+ to positive signal (typically the red wire) and IN- to the negative signal (typically the black wire). Fig. A Fig. A  Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the screw in the back. Remove the battery from the battery sockets. Replace the battery with a Lithium U9VL-J. Close the case and secure with the screw. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. PHONETEMP The PHONETEMP differs from other MadgeTech data loggers by communicating through an onboard modem. In order to communicate to the PHONETEMP, the data logger must be connected to a phone line. In order to use the software to connect to the data logger, the correct COM port must be selected in our software. If you are unsure which COM port the modem is on, you can use Windows Device Manager to locate the port number. The PHONETEMP communicates at 2400 baud; therefore, the correct baud rate in the software must be selected. To connect, use the Modem Connection feature of the MadgeTech software located under the communication menu. In the telephone number field, enter the telephone number of the line that the PHONETEMP is on. The PHONETEMP will then answer the call and connect to the MadgeTech software. To start the PHONETEMP refer to HYPERLINK \l "Quick_Start"Quick Start and skip ahead to section 4. After starting, the PHONETEMP is armed; data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the screw in the back. Remove the battery from the battery sockets. Replace the battery with a Lithium U9VL-J. Close the case and secure with the screw. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. IROCTTEMP To start the IROCTTEMP refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the IROCTTEMP is armed; data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. The IROCTTEMP has eight SMP connections which allows the user to insert subminiature thermocouple plugs into the connectors on the device. Fig. A Fig. A  The software accepts eight different thermocouple types: J, K, T, E, R, S, B and N. To select the correct thermocouple type in the MadgeTech software to match the thermocouple that is being used: Select Thermocouple Type from the device menu. Fig. B Click on the change button in the Thermocouple Type window. Select the correct thermocouple type from the drop down list. Click on the Save button to store the thermocouple type in the device. Click Ok, to finish. Fig. B Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the four screws on the top of the case. Disconnect the battery from the battery connector. Replace the battery with a Lithium U9VL-J. Close the case and secure with screws, making sure that the wires for the battery connector do not get pinched between the cover and the case. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. IRQUADTEMP To start the IRQUADTEMP refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the IRQUADTEMP is armed; data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. The IRQUADTTEMP has four SMP connections which allows the user to insert subminiature thermocouple plugs into the connectors on the device. Fig. A Fig. A The software accepts eight different thermocouple types: J, K, T, E, R, S, B and N. To select the correct thermocouple type in the MadgeTech software to match the thermocouple that is being used: Select Thermocouple Type from the device menu. Fig. B Click on the change button in the Thermocouple Type window. Select the correct thermocouple type from the drop down list. Click on the Save button to store the thermocouple type in the device. Click Ok, to finish. Fig. B Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the four screws on the top of the case. Disconnect the battery from the battery connector. Replace the battery with a Lithium U9VL-J. Close the case and secure with screws, making sure that the wires for the battery connector do not get pinched between the cover and the case. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. IRTC110 To start the IRTC110 refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the IRTC110 is armed; data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. This device communicates at 57,600 baud. If you do not have the software set to Always Auto Configure, you must select this baud rate from the communication menu. The IRTC110 has a three position removable screw terminal. This allows the IRTC110 to be connected to most 2-wire thermocouples with lead wires, and the ability to shield the thermocouple with the ground connection. Be sure to connect the thermocouple with the right polarity as marked on the case of the device, the ground connection is optional. Fig. A Fig. A The software accepts eight different thermocouple types: J, K, T, E, R, S, B and N. To select the correct thermocouple type in the MadgeTech software to match the thermocouple that is being used: Select Thermocouple Type from the device menu. Fig. B Click on the change button in the Thermocouple Type window. Select the correct thermocouple type from the drop down list. Click on the Save button to store the thermocouple type in the device. Click Ok, to finish. Fig. B  Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the screw in the back. Remove the battery from the battery sockets. If necessary, trim the new battery leads to match the length of the existing battery leads. Replace the battery with a Lithium LTC-7PN. Close the case and secure with the screw. For 10+ year operation, be sure to set the reading rate to no more than one reading every 15 minutes. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. IRTC4000 To start the IRTC4000 refer to HYPERLINK \l "Quick_Start"Quick Start. After starting, the IRTC4000 is armed; data will start recording. The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer. Depending on the physical configuration of the IRTC4000, it will either have an input for a subminiature thermocouple connector or a removable screw terminal. The standard connector for the IRTC4000 is the SMP connection which allows for the user to insert the subminiature thermocouple plug into the connector on the device. The non-standard connector for the IRTC4000 is the removable screw terminal. This allows the IRTC4000 to be connected to most 2-wire thermocouples with lead wires. Be sure to connect the thermocouple with the right polarity as marked on the case of the device. Fig. A Fig. A The software accepts eight different thermocouple types: J, K, T, E, R, S, B and N. To select the correct thermocouple type in the MadgeTech software to match the thermocouple that is being used: Select Thermocouple Type from the device menu. Fig. B Click on the change button in the Thermocouple Type window. Select the correct thermocouple type from the drop down list. Click on the Save button to store the thermocouple type in the device. Click Ok, to finish. Fig. B  Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. Battery Replacement Open the case by removing the screw in the back. Remove the battery from the battery sockets. If necessary, trim the new battery leads to match the length of the existing battery leads. Replace the battery with a Lithium LTC-7PN. Close the case and secure with the screw. Note: Before replacing the battery, be sure to stop and download the data from the logger. This will ensure that the date and time stamp of the data set is accurate. BRIDGE110 Battery Warning MadgeTech data loggers contain a lithium battery. Do not cut open, incinerate, heat above 85 C, recharge, submerge under water or other liquid, or expose the battery to rain or excess humidity. Dispose per local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. ENGINEERING UNITS To use the programmable engineering units for applicable devices, select Engineering Units from the device menu in the software. There are two options for entering the engineering units. The first option is to use the Engineering Units Wizard. This method will automatically calculate the gain and offset values based on the information provided. To start the wizard, click on the Wizard button in the Engineering Units window and follow these simple instructions: The Engineering Units Wizard window will appear. The first screen is an introduction to the Wizard, click next to continue. Enter the label of the units that are to be used. Click next to continue. Enter the description of the units to be used. Click next to continue. Enter the range in which the device will be measuring. If the device will read 0-100 mA, but only 4-20 mA is going to be used, enter a low limit of 4 and a high limit of 20. Click next to continue. Enter the range in which you want the data to be displayed. For example if you want to see 0-14 pH, then enter 0 in the low limit box, and enter 14 in the high limit box. Click next to continue. Click finish to return to the Engineering Units screen. When finished, click save to store the engineering units in the device. After the software has downloaded all the data, click ok to exit out of the Engineering Units setup. The second option is to calculate the gain and offset by hand and enter those values in the Engineering units gain and offset fields. Engineering units consist of two parts, the gain and the offset. In order to calculate the gain and offset, look at the example below. Here we are trying to convert mA on a scale of 4 to 20 to pH on a scale of 0 to 14. Gain = 14 pH 0 pH=14 pH=0.87520 mA 4 mA16 mA Offset = 14 pH Gain * 20 mA=-3.5 pH After calculating the gain and offset, enter those values in their respective fields. You will also need to enter a unit description and label in their respective fields. When finished, click save to store the engineering units in the device. After the software has downloaded all the data, click ok to exit out of the Engineering Units setup. Calibration MadgeTechs software allows for our data loggers to be calibrated by the user without mechanical calibration. To calibrate MadgeTech devices, simply compare the readings taken by the device to actual known parameters. To open the calibration window, select Calibration under the device menu in the software. There are two options in setting the calibration values into the device. The first option is to use the Calibration Wizard. This method will automatically calculate the gain and offset values based on the information provided. To start the wizard, click on the Wizard button in the Calibration window and follow these simple instructions: The Calibration Wizard window will appear. The first screen is an introduction to the Wizard, click next to continue. Depending on the device to be calibrated, a channel may require a one or two point calibration. If the channel requires a one point calibration, then enter the actual known value in the Calibration Point field. Enter the value read by the device in the Value Read by the Device at the Calibration Point field. Click next to continue. If the channel requires a two point calibration, there will be two sets of two fields. One set will be the Low Measurement and Reading (LMR), the other set will be the High Measurement and Reading (HMR). In the LMR section, enter the actual known value in the top box. Enter the value read by the device in the Reading field. In the HMR section, enter the actual known value in the top box. Enter the value read by the device in the Reading field. Repeat step 2 for each channel the device has. When finished, click finish to return to the Calibration screen. When finished, click save to store the calibration values in the device. After the software has downloaded all the data, click ok to exit out of the calibration setup